Related Experiment Videos
[Comparison of corneal topography and keratometer in patients with cataract preoperatively and postoperatively]
1Zhongshan Ophthalmic Center, Sun Yat-sen University of Medical Sciences, Guangzhou, China.
Abstract:
The comparison of corneal topography with keratometer in 200 patients (278 eyes) was made. The results revealed that Sim K values of corneal topography were 44.22D and 44.26D, mean diopters of keratometer were 44.16D and 44.23D in 150 eyes of 116 patients with cataract preoperatively. There was no significant difference between Sim K values of topography and diopters of keratometer (P > 0.05). The Sim K values of the topography were 46.37D and 43.06D and diopters of keratometer were 46.09D and 42.88D in 90 eyes of 67 patients who received cataract extraction with IOL implantation or second IOL implantation. No significance was found in results between Sim K values and diopters of keratometer in patients postoperatively (P > 0.05). Our investigation demonstrated that Sim K values of corneal topography were in agreements with diopters of keratometer in patients with cataracts preoperatively or postoperatively. However, corneal topography has great values in diagnosis of corneal diseases, traumatic astigmatism, severe non-rule stigmatisms, in which keratometer is often useless.