Rapid column liquid chromatographic analysis of flucloxacillin in plasma on a microparticulate pre-column
1Department of Pharmacy, University of Queensland, Brisbane, Australia.
Abstract:
A rapid, sensitive, accurate and precise high-performance liquid chromatographic assay is described for flucloxacillin in plasma. Samples (100 microliters) containing dicloxacillin (internal standard) were extracted with ethyl acetate (2 ml). The mobile phase of acetonitrile (18%, v/v) in phosphate buffer (0.01 M, pH 7) was pumped at 1.2 ml/min through a 40 x 3.2 mm I.D. column (3 microns particles). Detection was at 220 nm. Calibration plots were linear (r > 0.9995) from 0.2 to 40 mg/l. Within-day and between-day coefficients of variation were less than 9% (n = 6). The detection limit was 0.05 mg/l and the limit of quantitation was 0.3 mg/l. Of 24 drugs tested, only phenytoin and carbamazepine may interfere in some patients' samples.