Serum xylose analysis by gas chromatography/mass spectrometry
J C Deutsch1, V R Kolli, C R Santhosh-Kumar
1Division of Gastroenterology, University of Colorado Health Sciences Center, Denver 80262.
Abstract:
A gas chromatography/mass spectrometric (GC/MS) isotope dilution assay for xylose was developed using tertbutyldimethylsilyl-derivatized xylose and [13C]1xylose, and applied to human serum samples. A calibration curve in serum using this assay showed < 3% variation (< 10 mg/L) for any given point. The correlation coefficient for xylose measurements made on 27 sera between a colorimetric method performed by a national commercial reference laboratory and the GC/MS method developed here was .952. However, xylose determinations of 10 of 27 samples differed by > 10% (up to 150 mg/L) when colorimetric values were compared to GC/MS. Two of these samples had borderline-low xylose values by GC/MS, but were well within the normal range by colorimetric analysis. gas chromatography/mass spectrometric isotope dilution assay appears to be an accurate method to measure xylose in serum. These data also suggest that further prospective studies comparing GC/MS to colorimetric methods are indicated for subjects undergoing oral xylose testing.