Disc failure in a Björk-Shiley mitral prosthesis
C Van Sabben1, S C Griffin, G E Venn
1Department of Cardiac Surgery, St Thomas' Hospital, London, United Kingdom.
The Journal of Heart Valve Disease
|September 1, 1994
Summary
A 50-year-old man
Area of Science:
- Cardiovascular Surgery
- Biomaterials Science
Background:
- Björk-Shiley spherical valves are mechanical heart valves used in cardiac surgery.
- Valve malfunction can necessitate reoperation, posing surgical risks.
- Pyrolytic carbon is a common material used in prosthetic heart valves due to its durability.
Observation:
- A 50-year-old male patient presented with a malfunctioning Björk-Shiley spherical valve.
- The valve had been implanted 14 years prior to the current presentation.
- Reoperation was performed to address the valve dysfunction.
Findings:
- Examination of the explanted Björk-Shiley valve revealed a critical finding: a missing segment of the pyrolytic carbon disc.
- Despite the structural defect, there was no evidence of embolization, meaning the detached carbon fragment did not travel to other parts of the body.
- The patient experienced a satisfactory recovery following the surgical intervention.
Implications:
- This case highlights a potential failure mode in long-term implanted Björk-Shiley spherical valves, specifically pyrolytic carbon disc fracture.
- The absence of embolization in this instance suggests that such fractures may not always lead to catastrophic embolic events.
- Further investigation into the long-term durability and failure mechanisms of mechanical heart valves is warranted to improve patient outcomes and device design.