A comparison of three electron planning algorithms for a 16 MeV electron beam
P M Ostwald1, P E Metcalfe, J W Denham
1Department of Radiation Oncology, Newcastle Mater Misericordiae Hospital, Waratah, NSW, Australia.
Purpose:
We report results of a comparison of three electron planning algorithms, an Age-Diffusion Pencil beam algorithm and two (2-D) and three dimensional (3-D) Hogstrom pencil beam algorithms, using simple 2 x 2 cm air and hard bone inhomogeneities and a complex anthropomorphic head and neck phantom.
Methods And Materials:
The simple inhomogeneities have variable dimensions outside the plane of calculation to test the effects of out of plane scattering on 2-D algorithms, compared with dose measured by film below the inhomogeneity in the dose fall-off range. Comparisons are also made of a parotid treatment field for 16 MeV electrons, and the dose measured by high sensitivity thermoluminescent dosimeters in the head and neck phantom.
Results:
Behind the simple inhomogeneities, the electron algorithms are found to underestimate the dose behind the air cavity by up to 40% and overestimated the dose behind bone by up to 30%. In the head phantom, the presence of inhomogeneities also presents problems for the algorithms, with overestimations of dose of up to 20% found behind bone-tissue interfaces, apparently due to shielding by high density bone. Overestimations of up to 17% are also found beside interfaces parallel to the beam. Underestimations of dose of up to 10% are found on the beam-side of interfaces, due to under-prediction of backscattered electrons. All three investigated algorithms underestimate the dose by up to 20% behind extreme surface curvature. One algorithm is found to underestimate the dose in the falloff region while another overestimates the dose around the 90% isodose.
Conclusion:
Clinicians should be aware of the limitations of their planning systems.