Electromagnetic stimulation of the auditory system: effects and side-effects
1Department of Physiology II, Karolinska Institute, Stockholm, Sweden.
Summary
Extracranial electromagnetic stimulation (EMS) can evoke auditory sensations in deaf patients, but its acoustic artifact poses a risk of permanent hearing loss due to intense impulse noise.
Area of Science:
- Neuroscience
- Audiology
- Biomedical Engineering
Background:
- Extracranial electromagnetic stimulation (EMS) is a non-invasive technique for activating nervous systems.
- EMS is utilized in neurology and otolaryngology for brain and facial nerve stimulation.
- The auditory system's response and potential risks of EMS require investigation.
Purpose of the Study:
- To evaluate the efficacy of EMS in eliciting auditory sensations in profoundly hearing-impaired individuals.
- To assess the safety and side effects of EMS on the auditory system in humans and animals.
- To characterize the acoustic artifact generated by EMS and its potential for causing hearing damage.
Main Methods:
- Non-invasive EMS was applied to 18 deaf or severely hard-of-hearing patients using a magnetic coil at various cranial positions.
- Auditory sensations were recorded, and thresholds of auditory sensation (TAS) were determined.
- Acoustic measurements of the magnetic coil acoustic artifact (MCAA) were performed in animal models and human cranial models.
Main Results:
- EMS successfully elicited auditory sensations in 14 out of 18 patients across different coil placements.
- Thresholds of auditory sensation (TAS) ranged from 20-50% of the maximum EMS level (2.0 Tesla).
- The MCAA generated by EMS reached levels up to 160 dB SPL, with significant energy in the 2-5 kHz range, posing a risk for hearing loss.
Conclusions:
- EMS can stimulate residual auditory pathways and evoke sensations in deaf individuals.
- The high-intensity acoustic artifact produced by EMS is a significant side effect, capable of causing permanent sensorineural hearing loss.
- Further research is needed to mitigate the risks associated with the MCAA during EMS application.