Size of cortical bone and relationship to bone mineral density assessed by quantitative computed tomography image
O Louis1, P Van den Winkel, P Covens
1Department of Radiology, Vrije Universiteit Brussel, Belgium.
Rationale And Objectives:
The accuracy of the measurement of the size of cortical bone on computed tomography (CT) images of human vertebrae was evaluated using an automated contour detection and segmentation procedure.
Methods:
Forty human lumbar vertebrae were scanned using 8-mm slices and an automated detection for definition of trabecular and cortical region of interest. The vertebrae were embedded in a polyester resin and 8-mm-thick midvertebral specimens were excised using a diamond circular saw. Contact radiographs of these specimens were performed and, after photograph magnification, the cortical area was measured using computerized planimetry.
Results:
Cortical area measured on CT images was highly correlated with the area measured by planimetry on the specimens (r = .91; P < .001) with, however, a systematic over-estimation. A significant relationship was found between density and width of the cortex (r = .56; P < .001).
Conclusions:
Computed tomography is able to assess the size of cortical bone in human vertebrae, but a threshold detection algorithm, as used in the current study, is not adequate to obtain the precise anatomic dimensions.