Precision, accuracy, and data acceptance criteria in biopharmaceutical analysis
1Department of Pharmacy and Pharmaceutics, Virginia Commonwealth University, Richmond 23298-0533.
Pharmaceutical Research
|October 1, 1993
Summary
Evaluating analytical methods requires rigorous accuracy and precision checks. Significance testing offers a more stringent accuracy assessment than fixed interval bias, ensuring reliable bioanalytical results.
Area of Science:
- Analytical Chemistry
- Bioanalytical Science
Background:
- Accuracy and precision are critical for analytical method validation.
- Quality control (QC) monitoring is essential for ensuring the reliability of bioanalytical results.
Purpose of the Study:
- To evaluate various accuracy, precision, and quality control (QC) testing approaches for bioanalytical methods.
- To compare the stringency of different QC acceptance criteria.
Main Methods:
- Applied fixed interval bias and significance testing for accuracy.
- Utilized fixed interval percentage relative standard deviation (%RSD) and analysis of variance (ANOVA) for precision.
- Assessed QC acceptance criteria including +/- 20% fixed range, 99% confidence interval, multiple rules, and range charts.
Main Results:
- Significance testing identified three methods as unacceptable for accuracy, while fixed interval bias accepted all seven.
- Precision assessment using unconfounded %RSD or ANOVA effectively evaluated between- and within-run variations.
- The fixed interval bias method for accuracy assessment was found to be more liberal than significance testing.
Conclusions:
- Significance testing provides a more conservative and reliable assessment of accuracy compared to fixed interval bias.
- ANOVA is effective for assessing precision by separating sources of variation in confounded data.
- Robust QC strategies are vital for validating bioanalytical data integrity.