Related Experiment Videos
[Safety testing of high frequency surgery neutral electrodes (measuring instrument for impedance and temperature
1Institut für Experimentalphysik, Angewandte Physik und Mikroelektronik, Universität Innsbruck.
Biomedizinische Technik. Biomedical Engineering
|January 1, 1993
Abstract:
A new safety tester device for surgical grounding plates is described. This device can measure impedance distribution and temperature increase under a standard load in accordance with the AA-MI-HF18 standard. Apart from relevant information about the grounding plates under test, possible risks for the patient can also be demonstrated.