Related Experiment Videos
Complex bioelectric impedance measurement system for the frequency range from 5 Hz to 1 MHz
1Department of Veterans Affairs Medical Center, Milwaukee, WI.
Annals of Biomedical Engineering
|March 1, 1993
Abstract:
Analytic techniques that have been successfully employed in materials science, and to a lesser extent in the study of biologic systems, have potential for improving the application of bioelectric impedance provided that both real and imaginary impedance components can be measured with sufficient accuracy over a given frequency range. Since biologic tissue, particularly animal tissue, is typically highly conductive, phase angles are small, making accurate measurements difficult. A practical four-terminal system employing commercial lock-in amplifiers is described and error sources and corrective techniques are discussed.