Deconvolution of detector size effect for small field measurement
Related Experiment Video
Updated: Nov 6, 2026

From Fast Fluorescence Imaging to Molecular Diffusion Law on Live Cell Membranes in a Commercial Microscope
Published on: October 9, 2014
P D Higgins1, C H Sibata, L Siskind
1University of Minnesota, Department of Therapeutic Radiology, Minneapolis 55455-0110, USA.
Abstract:
Parametrization of the small fields employed in stereotactic applications is a painstaking process involving extensive film dosimetry to achieve acceptable beam edge definition. Use of cylindrical or spherical detectors for profile measurements would simplify data acquisition but add a volume averaging artifact to beam edge definition. We demonstrate a simple approach to unfolding the chamber size artifact from measured small beam profiles using typical cylindrical chambers. In comparison with film measurements we have found good agreement when the detector response function is deconvoluted from the measured profiles, although the amount of correction needed is fairly minimal for the detectors studied.