Theoretical considerations and in vitro results for the development of percutaneous transcatheter balloon embolectomy
J Brossmann1, T Jahnke, L R Frank
1Klinik für Radiologische Diagnostik, Christian-Albrechts-Universität Kiel, Germany.
Rationale And Objectives:
To prove the feasibility of performing percutaneous transcatheter embolectomy using a funnel-shaped catheter tip and balloon embolectomy catheters.
Methods:
A theoretical model is described to assess the influence of the diameter of an introductory device and coaxially introduced embolectomy catheter. A funnel-shaped catheter tip was attached to 7F, 8F, and 9F introductory sheaths. In vitro embolectomy of 5 g and 12 g thrombi was performed with 3F and 4F embolectomy catheters.
Results:
The number of extractions required was significantly related to the ratio of the diameters of the sheaths and embolectomy catheter shafts. The combination of a 7F sheath with a 4F embolectomy catheter required the greatest number of extractions (8.4 +/- 1.7). The least number of extractions was needed for the combination of a 9F sheath with 4F and 3F embolectomy catheters (1.6 +/- 0.7 and 1.8 +/- 0.4, respectively).
Conclusions:
Our results indicate that percutaneous embolectomy with balloon embolectomy catheters is feasible. However, further research is necessary before the final catheter design is chosen.