Related Experiment Videos
Carmeda surface heparinization in neonatal ECMO systems: long-term experiments in a sheep model
1Department of Pediatric Anesthesiology, St Göran's Hospital, Karolinska Institute, Stockholm, Sweden.
Perfusion
|September 1, 1995
Summary
Carmeda heparin coating significantly improves the thromboresistance of neonatal extracorporeal membrane oxygenation (ECMO) systems. This coating reduces clotting, potentially decreasing the need for systemic heparinization during neonatal ECMO.
Area of Science:
- Biomaterials Science
- Cardiovascular Engineering
- Neonatal Intensive Care
Background:
- Extracorporeal membrane oxygenation (ECMO) is a life-support technology for neonates with severe cardiorespiratory failure.
- Systemic heparinization is crucial for preventing circuit thrombosis but carries risks of bleeding.
- Developing thromboresistant biomaterials is essential to improve ECMO safety and efficacy.
Purpose of the Study:
- To evaluate the thromboresistance of Carmeda heparin-coated neonatal ECMO circuits compared to non-coated circuits.
- To assess the impact of heparin coating on clotting activation and fibrin monomer formation during ECMO.
- To determine the potential of heparin coating to reduce systemic heparinization requirements in neonatal ECMO.
Main Methods:
- A sheep model was used to compare three Carmeda heparin-coated ECMO systems (45-96 hours runtime) with three non-coated systems (12-66 hours runtime).
- Standardized flow rates (200 ml/min) and activated clotting time (ACT ≈ 120 seconds) were maintained.
- Clotting assessment and detection of fibrin monomers were performed at the end of experiments.
Main Results:
- Heparin-coated systems exhibited minimal clotting, whereas non-coated systems showed significant clotting throughout.
- Fibrin monomers were detected after 24 hours in the heparin-coated group but within 60 minutes in the non-coated group.
- The Carmeda heparin coating demonstrated a clear thromboresistant effect.
Conclusions:
- Carmeda heparin coating provides significant thromboresistance for neonatal ECMO circuits.
- Heparin coating effectively reduces thrombus formation and activation of the coagulation cascade.
- This technology may reduce the reliance on systemic heparinization in neonatal ECMO, enhancing patient safety.