Variation in output power of laser prostatectomy fibers: a need for power measurements
Related Experiment Videos
C F Van Swol1, E te Slaa, R M Verdaasdonk
1Department of Urology, Medical Laser Center, University Hospital Utrecht, The Netherlands.
Objectives:
The aim of this study was the assessment of the quality of side-firing fibers that are being used for laser prostatectomy, either by a laser light transmission measurement or by visual inspection.
Methods:
A power meter (Aquarius) was developed to measure the actual power transmitted through a side-firing fiber and delivered to the prostatic tissue. The power measurements were performed under clinical conditions, that is, under water and at relatively high input power. Furthermore, a protocol was developed for visual inspection of the fibers. Eight types of side-firing fibers were measured before use. Before and after a procedure, three fiber types were measured: ProLase II (28 samples), UltraLine (23 samples), and UroLase (44 samples). All these fibers were used in standard treatment protocols.
Results:
At 60 W the transmission of new fibers (not used) ranged between 49% and 83% when compared to a bare fiber. After use, a large variation was found in transmitted power between different samples of one device. A correlation with total transmitted power was not present. At higher power input, vapor bubbles are generated at the tip of the fibers. Depending on the fiber design, these bubbles have a major impact on the transmission. Only for the UroLase fiber was there a significant correlation between visual inspection and the transmission of used samples at 10, 20, and 40 W.
Conclusions:
The transmission strongly varies between fibers and between different samples of one fiber during clinical use. Moreover, the transmission does not correlate with visual inspection. A power measurement during a clinical treatment will contribute to a more controlled procedure and to a better comparison of clinical laser prostatectomy studies.