Clinical experience with the Nikkiso centrifugal pump
Y Shomura1, T Shimono, K Onoda
1Department of Thoracic and Cardiovascular Surgery, Mie University School of Medicine, Japan.
Artificial Organs
|June 1, 1996
Summary
The Nikkiso HPM-15 centrifugal pump offers safe and easy operation for cardiopulmonary bypass (CPB) and percutaneous cardiopulmonary support (PCPS). Clinical use demonstrated stable hemodynamics and minimal adverse effects, confirming its utility in critical cardiac procedures.
Area of Science:
- Cardiovascular Surgery
- Biomedical Engineering
- Medical Devices
Background:
- Preliminary data on the Nikkiso HPM-15 centrifugal pump for cardiopulmonary bypass (CPB) have been previously published.
- Recent clinical applications involved a newly developed controller for enhanced performance.
- This study details expanded clinical experiences with the Nikkiso HPM-15 pump.
Purpose of the Study:
- To evaluate the clinical safety and efficacy of the Nikkiso HPM-15 centrifugal pump.
- To assess the pump's performance in cardiopulmonary bypass (CPB) and percutaneous cardiopulmonary support (PCPS) procedures.
- To report on the ease of use and hemodynamic stability provided by the device.
Main Methods:
- The Nikkiso HPM-15 centrifugal pump was utilized in 23 cases.
- Twenty-two patients underwent CPB, and one patient with fulminant viral myocarditis received PCPS.
- Circuit preparation, deaeration, hemodynamic parameters, serum-free hemoglobin, platelet counts, and renal/liver function were monitored.
Main Results:
- The pump facilitated easy circuit preparation and effective deaeration.
- Stable hemodynamics were maintained throughout CPB and PCPS procedures in all patients.
- Minimal increases in serum-free hemoglobin, transient decreases in platelet count (recovering within 7 days), and high urine output were observed.
- No renal or liver function abnormalities were detected during CPB.
Conclusions:
- The Nikkiso HPM-15 centrifugal pump is safe and user-friendly.
- The pump demonstrates significant utility for both CPB and PCPS applications.
- Clinical outcomes suggest the device is a reliable option for extracorporeal circulatory support.