Single injection thermodilution. A flow-corrected method
J R Jansen1, J J Schreuder, J J Settels
1Department of Pulmonary Diseases, Erasmus University, Rotterdam, The Netherlands.
Anesthesiology
|September 1, 1996
Summary
A modified thermodilution method improves cardiac output estimation during controlled ventilation by correcting for blood flow variations. This flow-corrected equation enhances accuracy, reducing variability and enabling more reliable patient monitoring.
Area of Science:
- Cardiovascular Physiology
- Medical Engineering
- Critical Care Medicine
Background:
- Thermodilution technique for cardiac output measurement assumes constant blood flow.
- Controlled ventilation can cause flow modulations, leading to significant errors in cardiac output estimation.
- Existing methods lack accuracy in dynamic physiological states.
Purpose of the Study:
- To develop and validate a modified Stewart-Hamilton equation for accurate cardiac output measurement.
- To correct for errors introduced by ventilation-induced blood flow variations.
- To improve the reliability of thermodilution cardiac output estimates in mechanically ventilated patients.
Main Methods:
- A flow-corrected equation was developed, incorporating relative blood flow changes.
- Relative flow was calculated using pulmonary artery pressure and pulse contour analysis.
- Measurements were performed in 16 patients undergoing coronary artery bypass surgery, comparing catheter tip transducer and fluid-filled systems.
- Cardiac output was referenced against four uncorrected thermodilution estimates.
Main Results:
- Flow correction significantly reduced variability in cardiac output estimates from 10.6% to 5.0% in Group A (P < 0.0001).
- Limits of agreement for cardiac output narrowed from +/- 1.06 L/min to +/- 0.5 L/min post-correction.
- Adequate pressure waveform quality for correction was achieved with catheter tip transducers, but not consistently with fluid-filled systems.
Conclusions:
- A single thermodilution cardiac output estimate using the flow-corrected equation is clinically feasible.
- This method requires a more complex computation and an additional pressure measurement, often already available.
- The technique offers potential for considerable reduction in patient fluid load.