Intrasound vibration testing in acute ankle injuries
M C Plewa1, R Otto, J Verbrugge
1Vincent Medical Center, Toledo, OH 43608, USA. mcplewa@msn.com
Summary
Intrasound vibration testing (IVT) showed low sensitivity for diagnosing acute ankle fractures, with only 39% accuracy in detecting fractures. While specific, IVT is not a reliable standalone diagnostic tool for ankle injuries.
Area of Science:
- Orthopedics
- Diagnostic Imaging
- Emergency Medicine
Background:
- Acute ankle injuries are common in emergency departments.
- Radiography is the standard diagnostic tool for suspected ankle fractures.
- Novel diagnostic methods are being explored to improve accuracy and efficiency.
Purpose of the Study:
- To evaluate the diagnostic accuracy of intrasound vibration testing (IVT) compared to plain radiography for acute ankle fractures.
- To determine the sensitivity, specificity, and predictive values of IVT in diagnosing ankle fractures.
Main Methods:
- A group-sequential, nonrandomized, double-blind, observational study.
- Included patients aged 10+ with acute ankle injuries undergoing radiography.
- IVT applied to malleoli; positive if pain caused withdrawal. Excluded injuries >24h, inadequate data, or positive IVT on uninjured ankle.
Main Results:
- Of 97 analyzed patients, 13 had fractures (9 lateral malleolar, 1 medial malleolar, 1 bimalleolar, 2 talar).
- IVT detected only 5 of 13 fractures (sensitivity 39%). Specificity was 83%.
- Positive predictive value was 26%; negative predictive value was 90%. Overall accuracy was 77%.
Conclusions:
- Intrasound vibration testing demonstrates low sensitivity and poor positive predictive value for diagnosing acute ankle fractures.
- IVT is not a reliable standalone diagnostic tool for acute ankle fractures.
- Further research is needed to explore the utility of IVT in specific patient populations or as an adjunct to radiography.