Related Experiment Video
Updated: Aug 14, 2026

A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Empirical dose rate and attenuation data for radionuclides in nuclear medicine
1Department of Diagnostic Physics, Royal Brisbane Hospital.
Abstract:
Empirical dose rate shielding design data for lead and concrete barriers are presented for the radio-nuclides Tc-99m, I-131, Ga-67, T1-201, and Mo-99 for calculating occupational radiation exposures in Australian nuclear medicine facilities. These data have shown that the radiation dose rates received from stored radio-pharmaceuticals can be substantially higher than those calculated using the radiation dose rate and attenuation data provided in most nuclear medicine texts. The primary causes of the increased radiation dose rates are forward scattering (all nuclides), radiation hardening by selective absorption of low energy photons in lead shielding (Ga-67, Mo-99 and I-131) and contamination by radio-nuclides emitting high energy photons (T1-200 in T1-201).
Related Concept Videos
Radioactive Decay and Radiometric Dating
Biological Effects of Radiation
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Isotopes and Radioisotopes
An isotope containing more...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Drug Dosing in Renal Diseases: Dose Adjustments Based on Drug Clearance and Elimination Rate Constant