Accuracy of jitter estimates using different filter settings on Visi-Pitch: a preliminary report
C K Perry1, D R Ingrisano, S R Scott
1Voice and Phonology Research Lab, University of Wyoming, Laramie, USA.
This study compared Visi-Pitch jitter measurements with hand measurements, finding small differences. Recommended Visi-Pitch filter settings are supported for accurate jitter analysis.
Area of Science:
- Speech science and acoustics
- Voice analysis technology
Background:
- Accurate measurement of voice perturbation, specifically jitter, is crucial for diagnosing voice disorders.
- The Visi-Pitch device offers automated jitter analysis, but its optimal settings require validation.
- Previous research suggested alternative filter settings for Visi-Pitch, potentially impacting jitter measurement accuracy.
Purpose of the Study:
- To evaluate the accuracy of Visi-Pitch jitter measurements using manufacturer-recommended filter settings.
- To compare Visi-Pitch derived jitter values with traditional hand-measured values.
- To determine if alternative filter settings offer superior jitter estimation compared to standard settings.
Main Methods:
- Sustained /a/ phonations were recorded from three subjects.
- Jitter values were automatically calculated using Visi-Pitch with four different filter settings.
- Acoustic signals were manually measured, and jitter was computed using Koike's formula for comparison.
Main Results:
- The difference between Visi-Pitch and hand-measured jitter was minimal across tested settings.
- Manufacturer-recommended filter settings yielded jitter values comparable to manual measurements.
- Alternative settings did not demonstrate a significant improvement in jitter estimation accuracy.
Conclusions:
- The Visi-Pitch device provides reliable jitter measurements with its default filter settings.
- Manufacturer recommendations for Visi-Pitch filter settings are validated for clinical use.
- There is no significant benefit to using alternative filter settings for jitter analysis with Visi-Pitch.
Related Concept Videos
Uncertainty in Measurement: Accuracy and Precision
Uncertainty in Measurement: Reading Instruments
IR Spectrum Peak Splitting: Symmetric vs Asymmetric Vibrations
Determination of Expected Frequency
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Sampling Methods: Overview
In analytical chemistry, the choice of sampling...