Related Experiment Videos
[Automated blood component collection with the MCS 3p]
1Abteilung für Transfusionsmedizin und Gerinnungsphysiologie, Philipps-Universität Marburg, Deutschland.
Summary
Two automated blood component collection protocols were evaluated. Both showed good red blood cell quality and acceptable leukocyte/platelet levels, though FFP platelet counts were high.
Area of Science:
- Transfusion Medicine
- Blood Component Collection
- Apheresis Technology
Background:
- Automated blood component collection systems are crucial for optimizing blood product yields.
- Evaluating new protocols is essential for improving efficiency and product quality in apheresis.
- The MCS 3p cell separator offers potential for advanced blood component collection.
Purpose of the Study:
- To evaluate two novel protocols for the Haemonetics MCS 3p cell separator for automated blood component collection.
- To assess the technical feasibility, product quality, and biochemical integrity of collected blood components.
Main Methods:
- Two protocols, D-RBCP and RBCP-F, were used for automated collection of Fresh Frozen Plasma (FFP) and Red Blood Cells (RBCs).
- The RBCP-F protocol incorporated in-line filtration of RBCs in a closed system.
- Biochemical parameters and cell counts of RBCs were analyzed over 49 days; FFP and RBC components were assessed post-collection.
Main Results:
- Both protocols demonstrated successful erythroplasmapheresis without technical issues or adverse reactions.
- Biochemical parameters of RBCs, including hemoglobin, 2,3-DPG, ATP, and LDH, remained stable and comparable or superior to standard preparations up to 49 days.
- Leukocyte and platelet counts in RBCs and FFP (D-RBCP) met German guidelines, although FFP platelet counts exceeded standard values in the RBCP-F protocol.
Conclusions:
- The evaluated protocols are technically sound for automated blood component collection, yielding high-quality RBCs.
- While leukocyte reduction in RBCs was adequate, FFP platelet levels require further optimization.
- Extracorporeal volumes were noted as potentially large and may require donor-specific adjustments.