Related Experiment Videos
Fatigue properties and stem subsidence in wire coil reinforced PMMA bone cement: a preliminary in vitro study
1Department of Biomedical Engineering, College of Engineering, University of Iowa, Iowa, City 52242, USA.
Abstract:
The longevity of the cemented total hip joint replacement depends on the integrity of the cement fixation with respect to the interfaces of the metal stem and bone, and stress/strain transmission and its distribution. A simple stainless steel wire coil was placed around a tapered stem in a simulated cement mantle to counteract radial- and hoop-stress and dynamic compression-compression tests were performed. The results showed that the wire coil reinforced cement group had a higher fatigue (30%) and strength (35%) up to 10(5) loading cycles. The amount of stem subsidence during fatigue tests of the wire reinforced group was significantly reduced (2.5 times) compared with the control group.