Pulse oximeter performance during desaturation and resaturation: a comparison of seven models
N S Trivedi1, A F Ghouri, E Lai
1Department of Anesthesiology, University of California at Irvine, Orange 92668, USA.
Study Objective:
To compare pulse oximeter performance during induced hypoxemia.
Design:
Prospective investigation in human volunteers.
Setting:
Laboratory facility at a university medical center.
Patients:
8 unanesthetized, healthy ASA physical status I volunteers.
Interventions:
We evaluated the accuracy and response times of seven popular pulse oximeters during induced hypoxemia. Arterial blood fractional oxygen saturation (SaO2) measurements were performed simultaneously and considered a gold standard.
Measurements And Main Results:
All oximeters were accurate (+/-2%) while subjects were breathing room air. During maximal hypoxemia (induced by breathing a FIO2 = 10% in nitrogen), large differences were noted between oxygen saturation as measured by pulse oximetry (SpO2) and SaO2 values, with pulse oximeters consistently underreporting SpO2 when actual SaO2 values were 75% or less. The Ohmeda 3740 (Ohmeda, Boulder, CO) using an ear probe was the first to detect desaturation (change in SpO2 > 3%) in 4 of 8 subjects (p < 0.05), and the Nellcor N200 reflectance oximeter (Nellcor, Inc., Pleasanton, CA) was first in 3 of 8 subjects (p < 0.05). During resaturation (after administering 100% oxygen), the Novametrix Oxypleth (Novametrix, Wallingford, CT) was significantly faster than other oximeters (p < 0.05) to return to baseline (SpO2 = 98%).
Conclusion:
Most models of oximeters tested performed well when hemoglobin oxygen saturation was high, but all were inaccurate when SaO2 was approximately 75%. During induced hypoxemia, there were significant differences in the response times of oximeters tested, with no model demonstrably superior to others in all measures of performance.
More Related Videos
06:57Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...