Hydroxyapatite coating of threaded pins enhances fixation
G Magyar1, S Toksvig-Larsen, A Moroni
1Department of Orthopedics, University Hospital, Lund, Sweden.
Summary
Hydroxyapatite (HA) coated screws significantly improved fixation strength for knee osteoarthritis treatment compared to standard screws. HA coating enhanced screw retention in the tibial diaphysis and reduced loosening in the metaphysis, indicating better long-term stability.
Area of Science:
- Orthopedics
- Biomaterials Science
- Medical Device Engineering
Background:
- Osteoarthritis of the knee is a degenerative joint disease often requiring surgical intervention.
- External fixation is a common orthopedic technique for stabilizing fractures and bone segments.
- Hydroxyapatite (HA) coating is a biomaterial used to enhance bone integration and implant stability.
Purpose of the Study:
- To compare the insertion and extraction torque forces of standard tapered screws versus hydroxyapatite (HA)-coated screws in patients undergoing hemicallotasis for medial knee osteoarthritis.
- To evaluate the effect of HA coating on screw fixation stability in the proximal tibial metaphysis and diaphysis.
Main Methods:
- A randomized study involving 19 patients with medial knee osteoarthritis treated with hemicallotasis.
- 76 external fixation screws were used: standard tapered screws and HA-coated screws.
- Insertion and extraction torque forces were measured for screws placed in the proximal tibial metaphysis and diaphysis.
Main Results:
- No significant difference in insertion torque forces between standard and HA-coated screws.
- Extraction torque forces were significantly higher for HA-coated screws in the diaphysis (p < 0.005).
- Significantly fewer HA-coated screws loosened in the metaphysis (1/18) compared to standard screws (18/18).
- Standard screws in the diaphysis lost approximately 40% of fixation, while HA-coated screws retained full fixation.
Conclusions:
- Hydroxyapatite coating significantly enhances the long-term fixation stability of external fixation screws in the tibial diaphysis.
- HA coating reduces screw loosening in the tibial metaphysis, improving overall implant stability in knee osteoarthritis treatment.