A B Swanson1, G de Groot Swanson
1Orthopaedic Research Department, Blodgett Memorial Medical Center, Grand Rapids, MI 49506, USA.
Abstract:
A double stemmed flexible hinge implant was designed in 1974 for use with the Mayo type resection arthroplasty of the first metatarsophalangeal joint. Press-fit titanium grommets were developed in 1985 to protect the implant midsection from sharp bony edges and shearing forces that can lead to implant abrasion or fracture, and particulate synovitis. The study of 90 first metatarsophalangeal joint implant arthroplasties performed with encircling press fit titanium grommets showed favorable bone response around the implant stems and the bone to grommet interface, with absence of complications relating to particulate reactivity, implant or grommet fracture. These findings suggest that the grommets effectively protect the implant to bone interface, improve implant durability, and prevent particulate synovitis.