Carbon cartridge standards for 125I and suggested applications
M A Tries1, J P Ring, G E Chabot
1University of Massachusetts Lowell, 01854, USA.
Abstract:
Carbon cartridge standards were prepared to assess the activity of 125I incident on, and adsorbed in, cartridge samples during air sampling. Each cartridge standard consisted of an 125I-spiked filter paper at a known depth, ranging from 0 to 19 mm, embedded in approximately 34 g of 20-30 mesh activated carbon contained within a 6.35 cm diameter by 2.22 cm deep metal cartridge with screened openings. The total counting efficiency values range from 17.8 to 20.8% for cartridges counted at 3.2 mm from a thin-crystal NaI(Tl) detector. The standards were analyzed using a front/back counting technique, and fitting functions were developed relating the front/back net counts ratio and counting efficiency to the 125I depth of burial. A method for determining sample activity that accounts for exponential radioiodine loading in cartridge samples is compared to a less complicated technique that assumes all the radioiodine is located at an equivalent depth of burial that is based on the sample front/back net counts ratio. In addition, methods are presented for determining airborne 125I activity for constant and variable concentrations. Variable concentrations are assumed to occur in a fume hood duct by one or more bulk releases as a result of iodinations that are performed during a given sampling interval. The two methods are shown to have maximum relative deviations ranging from -16 to +16%.