X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Charge-coupled device panoramic radiography: effect of beam energy on radiation exposure
T T Farman1, A G Farman, M S Kelly
1School of Dentistry, University of Louisville, Kentucky 40292, USA.
Objective:
To determine the consequences for entrance exposure of varying the beam energy and operating current for panoramic radiography using a charge-coupled device receptor.
Material And Methods:
Images were made of a RANDO average man phantom (Alderson Research Laboratories, Stamford, CT) at kVcp settings of 60, 66, 70 and 80 and an mA of 2.0, 3.2, 6.4 and 10.0. The exposure cycle was set as recommended by the manufacturer at 17.6 s. Diagnostic image quality was rated by a panel of two oral and maxillofacial radiologists and one oral and maxillofacial pathologist. Entrance exposures were assessed using a 3 cc ionization chamber placed at the beam entry points while imaging the molar, premolar, and anterior teeth both using the DigiPan (Trophy Radiologie, Vincennes, France) CCD receptor and conventional T-Mat G film/Lanex Regular screens (Eastman Kodak, Rochester, NY, USA).
Results:
Acceptable image quality was attained with combinations of 60 kVcp and 3.2, 6.4 or 10 mA, 70 kVcp and 2.0, 3.2 or 6.4 mA; at 80 kVcp irrespective of the mA it was unacceptable. The maximum reduction in entrance dose was 77%, averaged over the three sites, at 70 kVcp and 2 mA.
Conclusion:
The DigiPan receptor produces satisfactory images with saving in entry exposure saving of approximately 70% when compared with a conventional film/rare earth screen combination.
More Related Videos
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Types of Radioactivity
Alpha (α) decay is the emission of an α particle from the nucleus. For example, polonium-210 undergoes α decay:
Biological Effects of Radiation
X-ray Imaging
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Absorption of Radiation
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...