Structural integrity of implanted WMT infinity hip and Osteonics Omnifit hip in fresh cadaveric femurs
T Q Lee1, D A Lewis, S L Barnett
1Department of PM&R, Long Beach Veterans Affairs Medical Center, California 90822, USA.
Abstract:
Five matched pairs of fresh human femurs were used to quantitatively assess the structural integrity of the implanted Wright Medical Technology modular Infinity hip and the Osteonics single-piece Omnifit hip. The results showed that neither bone implant system was able to recreate the femoral anteversion of the intact femur (p < 0.05). Although the micromotion at the bone-implant interface was well within the limits of achieving bone ingrowth, the axial micromotion was greater for the Infinity hip than for the Omnifit hip (p < 0.05), but the rotational micromotion did not show a significant difference between the two bone implant systems (p > 0.5). The anterior proximal femur deformation patterns were similar for both the two bone implant systems and intact femurs. However, the medial proximal femur deformation patterns showed that the Infinity hip-implanted femur was more similar to the intact femur than the Osteonics Omnifit hip-implanted femur.