Related Experiment Video
Updated: Aug 15, 2026

Oral Biofilm Formation on Different Materials for Dental Implants
Published on: June 24, 2018
Retention of cemented implant-supported restorations
D K Kent1, S Koka, M L Froeschle
1Department of Dental Practice Management, University of Nebraska Medical Center College of Dentistry, Lincoln 68583-0740, USA.
Purpose:
This investigation compared cement failure loads of a gold cylinder luted to a CeraOne titanium abutment (5-mm-high chimney) 24 hours postcementation when either zinc phosphate, Tempbond NE, or Tempbond luting agent was used.
Materials And Methods:
Luting agent (0.01 mL) was applied to the internal aspect of the gold cylinder, and the cylinder was seated with finger pressure followed by a load of 5 kg for 10 minutes. The specimens were stored in 100% humidity at 37 degrees C for 24 hours before a uniaxial tensile force was applied to the gold cylinder by a mechanical testing instrument. A load-displacement chart was plotted and the cement failure load (CFL) calculated in kilograms. CFLs were determined for each luting agent with the access opening for the CeraOne gold screw on the internal of the abutment either unfilled or filled with autopolymerizing resin.
Results:
A two-way ANOVA and Turkey's HSD revealed that zinc phosphate resulted in significantly higher (p < .05) mean CFLs than Tempbond or Tempbond NE. In general, filling the access opening did not affect mean CFLs for zinc phosphate and Tempbond NE. However, mean CFLs for gold cylinders cemented with Tempbond and a filled access opening were significantly higher than mean CFLs with an unfilled access opening. However, neither the filled nor the unfilled Tempbond group mean CFL was significantly different from either Tempbond NE mean CFL.
Conclusions:
Twenty-four hours postcementation, zinc phosphate luting cement significantly enhanced CFLs of a CeraOne gold cylinder luted to a 5-mm CeraOne titanium abutment compared with Tempbond and Tempbond NE. Filling the gold screw access opening of the abutment with autopolymerizing resin only led to higher mean CFLs when Tempbond was used as the luting agent. Filling the access opening had no effect (p > .05) on mean CFLs when Tempbond NE or zinc phosphate was used.