Related Experiment Videos
An efficient chromatographic system for lipoprotein fractionation using whole plasma
W Innis-Whitehouse1, X Li, W V Brown
1Department of Medicine, Emory University School of Medicine, Atlanta, GA 30322, USA.
Journal of Lipid Research
|April 21, 1998
Summary
This study validates a semi-automatic Fast Phase Liquid Chromatography (FPLC) method for isolating plasma lipoproteins. The validated FPLC procedure accurately quantifies cholesterol and triglycerides in low-density (LDL) and high-density lipoprotein (HDL) fractions.
Area of Science:
- Clinical Chemistry
- Biochemistry
- Analytical Chemistry
Background:
- Accurate quantification of plasma lipoproteins is crucial for cardiovascular risk assessment.
- Existing methods for lipoprotein isolation and quantification can be time-consuming and require large sample volumes.
Purpose of the Study:
- To validate a semi-automatic Fast Phase Liquid Chromatography (FPLC) procedure for efficient isolation of plasma lipoproteins.
- To establish modified enzymatic assays for determining low concentrations of triglycerides and cholesterol.
- To compare FPLC-based quantification with established methods like dextran sulfate/MgCl2 precipitation and ultracentrifugation.
Main Methods:
- Plasma lipoproteins were isolated using a semi-automatic FPLC system with a 200 microl injection volume.
- Modified enzymatic assays were employed for triglyceride and cholesterol determination using a Beckman CX-5 Autoanalyzer.
- FPLC results for cholesterol and triglyceride levels in low-density lipoprotein (LDL) and high-density lipoprotein (HDL) fractions were correlated with precipitation and ultracentrifugation methods.
Main Results:
- The FPLC method demonstrated high correlation with established methods for HDL cholesterol (r2 = 0.997) and LDL cholesterol (r2 = 0.988).
- Linear detection ranges were established for HDL cholesterol (18-95 mg/dl) and LDL cholesterol (41-435 mg/dl).
- Low fractional standard deviations were observed for quadruplicate runs across a wide range of lipoprotein concentrations.
Conclusions:
- The validated semi-automatic FPLC procedure offers an efficient and accurate method for plasma lipoprotein isolation and quantification.
- This method is suitable for determining low concentrations of triglycerides and cholesterol in LDL and HDL fractions.
- The FPLC approach provides a reliable alternative to traditional methods, with high precision and accuracy.