Evaluation of aerosol drug output from the OptiChamber and AeroChamber spacers in a model system
Related Experiment Video
Updated: Oct 29, 2026

A System to Create Stable Nanoparticle Aerosols from Nanopowders
Published on: July 26, 2016
R N Dalby1, S Somaraju, V S Chavan
1Department of Pharmaceutical Sciences, University of Maryland at Baltimore, School of Pharmacy, USA.
Abstract:
Metered-dose inhalers (MDIs) are an effective means of generating drug-containing aerosols targeted for delivery to intrapulmonary airways. Many problems associated with incorrect patient use of MDIs are mitigated by adding a valved spacer device to the inhaler mouthpiece. This in vitro study compared the efficiency of drug output through a new spacer device, OptiChamber (HealthScan Products Inc., Cedar Grove, NJ), to that of a device commercially available since the 1980s, AeroChamber (Monaghan Medical, Plattsburgh, NY). Testing utilized MDI formulations of albuterol, beclomethasone dipropionate, and cromolyn sodium. OptiChamber equaled or, in the majority of cases, exceeded AeroChamber in output of the three drugs at two simulated inspiratory flow rates. Drug output from OptiChamber was found to be less sensitive to changes in flow rate than that from AeroChamber. OptiChamber also showed less decrease in drug output than AeroChamber when time delays were introduced between MDI actuation and the start of a simulated inhalation. Mass median aerodynamic diameters of drugs exiting the two spacers were generally similar to those of drugs exiting the MDI alone. However, spacers were shown to nearly eliminate the output of large-size drug particles (>5.8 microm), which can result in oropharyngeal drug deposition. Emitted fine-particle drug (<5.8 microm) doses from OptiChamber were greater than those from AeroChamber with or without a delay between canister actuation and the start of a simulated inhalation. The results suggest that OptiChamber may provide more efficacious aerosol drug delivery than AeroChamber under both ideal and suboptimal conditions.
More Related Videos
05:07Development of a Nose-only Inhalation Toxicity Test Chamber That Provides Four Exposure Concentrations of Nano-sized Particles
Published on: March 18, 2019
07:28Dry Powder and Nebulized Aerosol Inhalation of Pharmaceuticals Delivered to Mice Using a Nose-only Exposure System
Published on: April 6, 2017