Bioelectric Analyses of an Osseointegrated Intelligent Implant Design System for Amputees
Published on: July 15, 2009
3D-FEA of osseointegration percentages and patterns on implant-bone interfacial stresses
G Papavasiliou1, P Kamposiora, S C Bayne
1Department of Prosthodontics, University of North Carolina at Chapel Hill, School of Dentistry 27599-7450, USA.
Objectives:
The degree of osseointegration and its patterns are important for the success of implants. 3D-FEA was used to determine interfacial stresses on a single tooth implant (IMZ) for four degrees of osseointegration (100, 75, 50 and 25%), and five patterns at 50% osseointegration (locally alternating, coronal only, apical only, facial only and lingual only).
Methods:
The implant was restored with a metal-ceramic crown and subjected to 10 MPa axial or oblique applied stress. Resolved stresses were examined at four heights along the implant-bone interface.
Results:
The degree of osseointegration did not affect resolved stress levels or distributions. Oblique loads elevated interfacial stresses 5 to 20 times. Stresses were always higher at the bone crest.
Conclusions:
Osseointegration patterns with crestal bone reduced both crestal and apical stresses. Apical only osseointegration produced much higher apical stresses. Crestal osseointegration and axial loads minimized overall stress.