Comparative quantification of gastric electrical activity and electrogastrograms

M P Mintchev1, K L Bowes

  • 1Department of Surgery, University of Alberta, Canada.

Medical & Biological Engineering & Computing
|June 6, 1998
PubMed
Summary

This study compared gastric electrical activity (GEA) monitoring using implanted versus combined implanted and cutaneous electrodes. Cutaneous recordings accurately reflected GEA, showing high sensitivity for diagnosing gastric motility disorders.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Real-time gastric motility monitoring using transcutaneous intraluminal impedance measurements (TIIM).

Physiological measurement·2014
Same author

Intraluminal ultrasonic probe for volumetric monitoring of liquid gastroesophageal reflux.

Physiological measurement·2012
Same author

Design, implementation and testing of an implantable impedance-based feedback-controlled neural gastric stimulator.

Physiological measurement·2011
Same author

Comparative gastric motility study of Enterra ™ Therapy and neural gastric electrical stimulation in an acute canine model.

Neurogastroenterology and motility·2010
Same author

Implantable functional gastrointestinal neurostimulation.

Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference·2009
Same author

Pilot acute study of feedback-controlled retrograde peristalsis invoked by neural gastric electrical stimulation.

Physiological measurement·2009

Area of Science:

  • Gastroenterology
  • Biomedical Engineering
  • Medical Devices

Background:

  • Gastric electrical activity (GEA) monitoring is crucial for diagnosing gastrointestinal motility disorders.
  • Assessing GEA typically involves invasive implanted electrodes.
  • Non-invasive methods for GEA assessment are needed to improve patient comfort and accessibility.

Observation:

  • This study quantitatively compared GEA variability using implanted serosal electrodes versus a combination of implanted and cutaneous electrodes.
  • Gastric electrical activity (GEA) was recorded from asymptomatic patients using serosal electrodes and from patients with both implanted and cutaneous electrodes.
  • Mean frequency (MF) values and probability density functions (pdfs) of GEA signals were analyzed.

Findings:

Related Experiment Videos

  • Normal GEA in asymptomatic patients exhibited a mean frequency of 2.5-3.75 cycles/min with a standard deviation of 0-0.45 cycles/min.
  • Bell-shaped pdfs with coinciding maxima indicated normal electrical coupling in GEA.
  • In a comparative 36-hour study, cutaneous electrodes reflected GEA with 94% sensitivity and 79% specificity.
  • Implications:

    • Cutaneous electrodes offer a promising non-invasive approach for monitoring gastric electrical activity.
    • This technology could lead to more accessible and patient-friendly diagnostics for gastric motility disorders.
    • Further research may validate the use of cutaneous electrodes in clinical settings for GEA assessment.