Related Experiment Video
Updated: Aug 12, 2026

Integration of Miniaturized Solid Phase Extraction and LC-MS/MS Detection of 3-Nitrotyrosine in Human Urine for Clinical Applications
Published on: July 14, 2017
Determination of 1-nitropyrene with enzyme-linked immunosorbent assay versus high-performance column switching
J Zühlke1, D Knopp, R Niessner
1Institute of Hydrochemistry, Technical University of Munich, Germany.
Abstract:
An enzyme-linked immunosorbent assay (ELISA) on microplates and a HPLC coupled-column switching method were compared for the determination of the nitroarene compound 1-nitropyrene in airborne particulate organic matter collected at a busy intersection over a period of 2 months. After purification of the sample extract with silica, in the multidimensional chromatographic method nitroarenes were separated on a RP18 precolumn from matrix constituents followed by on-line reduction to corresponding aminoarenes with a Pt catalyst on alumina and a further separation of 1-aminopyrene on a second RP18 column. Methanol-water (70:30, v/v) was the mobile phase used. With ELISA, a six-fold overestimation was obtained for untreated samples. After clean-up it was lowered to approximately 1.6-fold overestimation which was mainly caused by cross-reaction of 2-nitropyrene and 2-nitrofluoranthene.
More Related Videos
07:38A General Method for Detecting Nitrosamide Formation in the In Vitro Metabolism of Nitrosamines by Cytochrome P450s
Published on: September 25, 2017
07:32Detection of 3-Nitrotyrosine in Atmospheric Environments via a High-performance Liquid Chromatography-electrochemical Detector System
Published on: January 30, 2019