Related Experiment Videos
Vascular anastomosis using a biodegradable device with a heat-shrinking sleeve: a preliminary report
W Bähr1, R Rosbander, R Gutwald
1Department of Oral and Maxillofacial Surgery, University Hospital Freiburg i. Br., Germany.
Purpose:
The objective of this study is to present a simple mechanical anastomosis system capable of uniting vessels of differing diameters and to evaluate its applicability.
Materials And Methods:
A new mechanical system for anastomosis of small vessels was used in 30 rabbits to join 100 sectioned carotid arteries and jugular veins. The system is resorbable and consists of a cuff and a shrinkable sleeve. To carry out the anastomosis, one vessel end is pulled through the cuff and everted over it. The second vessel end is then slipped over the everted vessel end and the cuff. A heat-shrinking sleeve is positioned over the site and shrunk to produce permanent fixation.
Results:
The overall patency rate of the anastomoses was 96.6% in the arteries and 93.3% in veins. Microscopically, the intima-to-intima contact sites were bridged by an endothelial layer in the first postoperative weeks.
Conclusion:
Advantages of the system include a short insertion time, simple instrumentation, a high patency rate, and resorbability. However, the method requires additional vessel length for eversion, and it cannot be used for end-to-side anastomoses.