Comparison between enzyme-linked immunosorbent assay and cytotoxic cross-match procedures for detecting IgG

B Susskind1, R H Kerman, R Nelson

  • 1Division of Immunology and Organ Transplantation, University of Texas Health Science Center, Houston, USA.

Transplantation
|January 12, 1999
PubMed

Insights

The Cross-Stat assay, an enzyme-linked immunosorbent assay method, shows high sensitivity and negative predictive value for detecting IgG anti-donor HLA antibodies compared to standard complement-dependent cytotoxicity cross-match (CDC XM) methods. Further clinical relevance testing is recommended.

Area of Science:

  • Transplantation immunology
  • Immunodiagnostics

Background:

  • Complement-dependent cytotoxicity cross-match (CDC XM) has limitations including complement requirement, need for viable cells, detection of non-HLA antibodies, and subjective scoring.
  • Cross-Stat, an enzyme-linked immunosorbent assay (ELISA) based cross-match, was developed to overcome these limitations in detecting IgG anti-donor HLA antibodies.

Purpose of the Study:

  • To compare the performance of the Cross-Stat assay with the standard anti-human globulin (AHG)-enhanced CDC XM procedure.
  • To evaluate the sensitivity, specificity, and predictive values of the Cross-Stat assay in detecting anti-donor HLA antibodies.

Main Methods:

  • A comparative study involving 524 sera from 230 transplant candidates and 51 cadaveric donors.
  • Evaluation of results obtained from Cross-Stat and AHG-enhanced CDC XM procedures.

Main Results:

  • A significant correlation (P<0.001) was observed between Cross-Stat and AHG-CDC IgG XM results.
  • Cross-Stat demonstrated 100% sensitivity, 93% specificity, 73% positive predictive value, and 100% negative predictive value relative to AHG-CDC XM.
  • Flow cytometry supported Cross-Stat results in cases of discrepancy, suggesting greater sensitivity.

Conclusions:

  • The Cross-Stat assay offers theoretical advantages over CDC XM for detecting IgG anti-donor HLA antibodies.
  • Despite a 15% indeterminate sample rate, the high sensitivity and NPV warrant further investigation into the clinical relevance of ELISA-based XM methods.
Abstract