[False positive of DUPAN-2 caused by IgM-human anti mouse antibody]

Masaki Abe1, Miyuki Hyoki, Ikurou Abe

  • 1Central Clinical Laboratory, Jikei University Hospital, Minato-ku, Tokyo 105-8471, Japan. ma_abe@jikei.ac.jp

Rinsho Byori. the Japanese Journal of Clinical Pathology
|February 7, 2013
PubMed

Insights

A false-positive result for the DUPAN-2 cancer marker was identified in a patient due to IgM-human anti-mouse antibody (HAMA) activity. This HAMA in the patient

Area of Science:

  • Clinical Chemistry
  • Immunology
  • Oncology Diagnostics

Background:

  • DUPAN-2 is a tumor-associated antigen used in cancer diagnostics.
  • Human anti-mouse antibody (HAMA) interference can cause inaccurate immunoassay results.
  • Understanding assay interferences is crucial for accurate clinical interpretation.

Observation:

  • A 40s female patient exhibited persistently elevated DUPAN-2 levels (up to 770 U/mL) unexplained by clinical or radiological findings.
  • Standard tests showed no dilution linearity, but absorption tests indicated interference primarily with IgM.
  • Specific immunological assays suggested patient's IgM, with HAMA activity, reacted with mouse monoclonal antibodies in the DUPAN-2 assay reagents.

Findings:

  • The elevated DUPAN-2 levels were confirmed as a false-positive phenomenon.
  • The mechanism involved the patient's IgM-HAMA cross-reacting with mouse monoclonal antibodies used in the DUPAN-2 assay.
  • High-performance liquid chromatography (HPLC) analysis showed the patient's DUPAN-2 eluted in the IgM fraction, unlike controls.

Implications:

  • This case highlights the potential for IgM-HAMA to cause false-positive DUPAN-2 results.
  • Clinicians should consider HAMA interference when DUPAN-2 levels are unexpectedly high and inconsistent with clinical presentation.
  • Further investigation into immunoassay interferences is necessary for reliable tumor marker testing.