Assessment of serum thyroxine binding capacity-dependent biases in free thyroxine assays
N D Christofides1, E Wilkinson, M Stoddart
1Research and Development, Ortho-Clinical Diagnostics, Cardiff Laboratories, Whitchurch, Forest Farm Estate, Cardiff CF4 7YT, Wales, UK.NCHRISTOPHIDES@compuserve.com
Background:
Free thyroxine (FT4) assays may exhibit biases that are related to serum T4 binding capacity (sBC). We describe two tests that can be used to assess the presence and magnitude of sBC-dependent biases in FT4 assays.
Methods:
We used a direct equilibrium dialysis FT4 assay as the reference method and compared the results obtained with those of the FT4 assays under investigation, in patient sera having a wide range of sBC. We then compared the expected and observed FT4 results for sera diluted with an inert buffer. Because serum dilution causes a predictable decrease in sBC, an increasingly negative bias on progressive dilution is indicative of a sBC-dependent bias.
Results:
The automated FT4 assay investigated (Vitros FT4) showed no demonstrable sBC-dependent bias by either test.
Conclusion:
These two tests can be used to screen for sBC-dependent biases in FT4 assays.
Related Concept Videos
The Equilibrium Binding Constant and Binding Strength
Factors Affecting Protein-Drug Binding: Patient-Related Factors
Age stands as a key determinant in protein-drug binding. Neonates, characterized by low albumin content, experience heightened concentrations of unbound drugs such as phenytoin and...
Physiological Pharmacokinetic Models: Assumption with Protein Binding
Graves Disease II: Pathophysiology