Intermittent P wave sensing in a patient with DDD pacemaker.
S Paraskevaidis1, S Mochlas, S Hadjimiltiadis
1First Cardiology Clinic, Aristotelian University, AHEPA Hospital, Thessaloniki, Greece.
Pacing and Clinical Electrophysiology : PACE
|May 11, 1999
Summary
Pacemaker T-wave oversensing can cause intermittent P-wave sensing issues in DDD devices. Decreasing ventricular sensitivity resolved this sensing problem in a patient with a DDD pacemaker.
Area of Science:
- Cardiology
- Biomedical Engineering
- Electrophysiology
Background:
- Dual-chamber pacemakers (DDD) are crucial for managing bradyarrhythmias.
- Accurate sensing of intrinsic cardiac electrical activity is vital for appropriate pacemaker function.
- T-wave oversensing is a known, albeit infrequent, complication in pacemaker therapy.
Observation:
- A patient with a DDD pacemaker experienced intermittent failure to sense P waves.
- The intermittent P wave sensing failure was correlated with T-wave oversensing by the ventricular lead.
- This oversensing erroneously triggered an extended atrial refractory period, masking subsequent P waves.
Findings:
- T-wave oversensing by the ventricular lead was identified as the cause of intermittent P-wave non-sensing.
- The pacemaker's programming led to an extended atrial refractory period, preventing detection of native P waves.
- Adjusting the ventricular sensitivity parameter successfully resolved the oversensing and restored P-wave sensing.
Implications:
- This case highlights the importance of considering T-wave oversensing in diagnosing pacemaker sensing abnormalities.
- Appropriate pacemaker programming, specifically sensitivity adjustments, can effectively manage oversensing issues.
- Optimizing pacemaker settings is essential for ensuring reliable cardiac rhythm management and patient outcomes.