Calibrated nanoinjections of mercury vapor
1University of Regensburg, Institute of Analytical Chemistry, Chemo- and Biosensors, Germany.
Abstract:
A simple technique for the calibration of mercury vapor chemosensors is described. It is based on a reductive deposition of a well-defined quantity of mercury (0) onto gold wire followed by thermal evaporation of the mercury into the sensor cell. The quantity of mercury on the gold wire was measured electrochemically by oxidation before and after heating, as well as after storage under different conditions. It is shown that the technique can be used for calibrated injections of nanogram quantities of mercury. The approach was applied to test the performance of ultrasensitive mercury chemoresistors.