Related Experiment Video
Updated: Aug 14, 2026

Treatment of Liver Metastases Using an Internal Target Volume Method for Stereotactic Body Radiotherapy
Published on: May 8, 2018
Dosimetry of 192Ir sources used for endovascular brachytherapy
N Reynaert1, M Van Eijkeren, Y Taeymans
1Department of Biomedical Physics, Faculty of Medicine, Ghent University, Gent, Belgium.
Abstract:
An in-phantom calibration technique for 192Ir sources used for endovascular brachytherapy is presented. Three different source lengths were investigated. The calibration was performed in a solid phantom using a Farmer-type ionization chamber at source to detector distances ranging from 1 cm to 5 cm. The dosimetry protocol for medium-energy x-rays extended with a volume-averaging correction factor was used to convert the chamber reading to dose to water. The air kerma strength of the sources was determined as well. EGS4 Monte Carlo calculations were performed to determine the depth dose distribution at distances ranging from 0.6 mm to 10 cm from the source centre. In this way we were able to convert the absolute dose rate at 1 cm distance to the reference point chosen at 2 mm distance. The Monte Carlo results were confirmed by radiochromic film measurements, performed with a double-exposure technique. The dwell times to deliver a dose of 14 Gy at the reference point were determined and compared with results given by the source supplier (CORDIS). They determined the dwell times from a Sievert integration technique based on the source activity. The results from both methods agreed to within 2% for the 12 sources that were evaluated. A Visual Basic routine that superimposes dose distributions, based on the Monte Carlo calculations and the in-phantom calibration, onto intravascular ultrasound images is presented. This routine can be used as an online treatment planning program.
Related Concept Videos
Biological Effects of Radiation
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET

