Multiple pulse generator malfunctions with a dual chamber pacemaker.

Igor Lipchenca1, David Barlev, David Luria

  • 1Heart Institute, Sheba Medical Center, Tel Hashomer, Israel.

Pacing and Clinical Electrophysiology : PACE
|July 26, 2002
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Leadless Pacemaker Implantation in High-Risk Patients: Real-World Outcomes From a Multicenter Israeli Registry.

Pacing and clinical electrophysiology : PACE·2026
Same author

Safety and performance of vibration-assisted extraction of chronic cardiac implantable electronic device leads: A pre-clinical ovine study.

Heart rhythm·2026
Same author

First-in-Human Mechanical Leaflet Modification of Native Aortic Stenosis During TAVR.

JACC. Case reports·2026
Same author

Modeling day-long ECG signals to predict heart failure risk with explainable AI.

NPJ digital medicine·2026
Same author

Impact of pulmonary hypertension on atrial fibrillation recurrence after pulmonary vein isolation: A prospective multicenter registry study.

International journal of cardiology. Heart & vasculature·2026
Same author

Subcutaneous Defibrillator Implantation With or Without Defibrillation Test: The Primary Results of the Randomized PRAETORIAN-DFT Trial.

Circulation·2026
Same journal

Dual Coronary Sinus Lead Strategy to Avoid Tricuspid Valve Traversal in Biventricular Pacing.

Pacing and clinical electrophysiology : PACE·2026
Same journal

A Case of Permanent Pacemaker Implantation via the Epicardial Approach Using the 3830 Lead in an 11-Day-Old Neonate (With Follow-Up of the Above Case).

Pacing and clinical electrophysiology : PACE·2026
Same journal

Cryoballoon Versus Radiofrequency Ablation for Persistent Atrial Fibrillation: Meta-Analysis of Randomized Trials.

Pacing and clinical electrophysiology : PACE·2026
Same journal

Tilt Test Duration in Suspected Vasovagal Syncope: Temporal Patterns and Diagnostic Yield in Patients From Central China.

Pacing and clinical electrophysiology : PACE·2026
Same journal

Combined Leadless Pacing and Subcutaneous ICD Therapy in Long QT Syndromes.

Pacing and clinical electrophysiology : PACE·2026
Same journal

Association of Anesthesia Modality With Procedural Parameters and Clinical Outcomes in PVI for Atrial Fibrillation.

Pacing and clinical electrophysiology : PACE·2026
See all related articles

The Sorin Biomedica Minidual 50 pacemaker experienced significant malfunctions, with over 10% failing completely. Researchers recommend replacing these devices due to unpredictable failure rates.

Area of Science:

  • Cardiology
  • Biomedical Engineering
  • Medical Device Evaluation

Background:

  • The Minidual 50 is a dual-chamber pacemaker.
  • Pacemaker reliability is crucial for patient health.
  • Previous data on the Minidual 50's long-term performance is limited.

Purpose of the Study:

  • To evaluate the clinical performance and reliability of the dual-chamber pacemaker Minidual 50.
  • To identify the incidence and characteristics of malfunctions associated with the Minidual 50.
  • To assess the long-term safety and efficacy of the Minidual 50 pacemaker.

Main Methods:

  • A retrospective analysis of 66 Minidual 50 pacemakers implanted between 1995 and 1998.
  • Follow-up duration averaged 33 months.
  • Kaplan-Meier survival analysis was used to predict malfunction-free survival.

Related Experiment Videos

Main Results:

  • Total function loss occurred in 10.6% of units.
  • False threshold measurements (sensing/pacing) were observed in 4.5% of patients.
  • Average time to malfunction was 37 months; failures were unpredictable and unrelated to battery status.
  • Predicted 4-year malfunction-free survival was 70%.

Conclusions:

  • The Minidual 50 pacemaker exhibits a high rate of unpredictable malfunction.
  • The study recommends replacing all remaining Minidual 50 units, especially in pacemaker-dependent patients.
  • Further investigation into the causes of Minidual 50 failure is warranted.