CR-39 acceptance test and optimisation for fast neutron dosimetry applications
E Fantuzzi1, B Morelli, G Falangi
1ENEA, Institute for Radiation Protection, Bologna, Italy. fantuzzi@bologna.enea.it
Radiation Protection Dosimetry
|October 18, 2002
Summary
Quality control of CR-39 plastic is crucial for reliable fast neutron dosimetry. Acceptance testing of CR-39 sheets ensures consistent performance in dosemeters used by the Individual Monitoring Service (IMS).
Area of Science:
- Nuclear Science and Engineering
- Radiation Detection and Measurement
- Materials Science
Background:
- The ENEA fast neutron dosemeter utilizes planar poly allyl diglycol carbonate (PADC) in a polyethylene holder.
- CR-39 plastic has been integral to the ENEA Individual Monitoring Service (IMS) for neutron dosimetry since 1998.
- Routine quality control involves acceptance tests for background track density and neutron sensitivity of new CR-39 batches.
Purpose of the Study:
- To analyze acceptance testing data for CR-39 plastic sheets used in ENEA's neutron dosemeters.
- To compare the dosimetric performance of standard CR-39 with CR-39 modified with dioctylphthalate (DOP).
- To evaluate the impact of CR-39 quality on dosemeter performance and lower detectable dose limits.
Main Methods:
- Analysis of acceptance testing data from over 30 CR-39 sheets produced between 1998 and 2000.
- Irradiation of dosemeters with a 241Am-Be neutron source at ENEA-IRP.
- Chemical etching procedure involving pre-etching and extended etching in KOH solutions.
Main Results:
- Demonstrated significant variability in background track density and neutron sensitivity among CR-39 batches.
- Identified that CR-39 composition, including the addition of DOP, affects background fluctuations and neutron sensitivity.
- Established a correlation between CR-39 plastic quality and the overall dosimetric performance of the fast neutron dosemeter.
Conclusions:
- Rigorous acceptance testing of CR-39 plastic is essential for ensuring the quality and reliability of fast neutron dosemeters.
- The intrinsic properties of CR-39, such as background track density and neutron sensitivity, directly impact dosimetry accuracy.
- Material quality control is paramount for maintaining the effectiveness of individual monitoring services.