R wave undersensing caused by an algorithm intended to enhance sensing specificity in an implantable cardioverter
Frank Eberhardt1, Werner Peters, Frank Bode
1Medizinische Klinik II, Universitätsklinik Luebeck, Luebeck, Germany.
Abstract:
This article reports on a case of ventricular undersensing despite normal R wave amplitudes during sinus rhythm in an ICD patient. Undersensing of ventricular signals was noted without any evidence of lead dislocation or variation in signal amplitude. Undersensing was due to an exceptionally small R wave signal width and a feature of the Biotronik sensing algorithm designed to avoid oversensing. This algorithm, intended to enhance the sensing specificity of the device, requires registration of two consecutive points above the maximum programmed sensitivity for a ventricular sense event. After modifying the algorithm to a single point registration undersensing disappeared.