Maximum angle of horizontal strabismus consistent with true stereopsis.
David A Leske1, Jonathan M Holmes
1Department of ophthalmology, Mayo Clinic, Rochester, MN 55905, USA.
Summary
True stereopsis may be rare in patients with horizontal strabismus greater than 4 prism diopters (PD). The Frisby test effectively identifies true stereopsis, unlike other stereoacuity tests with monocular clues.
Area of Science:
- Ophthalmology
- Vision Science
- Clinical Research
Background:
- Strabismus significantly impacts visual development and function.
- Accurate assessment of stereopsis in strabismus is crucial for treatment planning.
Purpose of the Study:
- To evaluate stereoacuity tests in patients with horizontal strabismus.
- Determine the maximum horizontal strabismus angle for true stereopsis.
- Assess the rate of false-positive results with different tests.
Main Methods:
- Prospective study of 186 patients (age ≥5) with constant horizontal strabismus.
- Stereoacuity measured using Titmus Fly, Animals, Circles, original Preschool Randot, and Frisby tests.
- Defined false-positive results (>20 prism diopters [PD]) and true-positive results based on established criteria.
Main Results:
- False-positive rates varied: Titmus Fly (6%), Animals (10%), Circles (35%), Randot (10%).
- No false-positive results occurred with the Frisby test.
- True stereopsis was absent in patients with >10 PD deviation; threshold may be as low as 4 PD.
Conclusions:
- Titmus tests (Fly, Animals, Circles) have limitations due to monocular clues.
- Frisby test is valuable for rapid presence/absence assessment of stereopsis.
- New Preschool Randot test quantifies stereopsis; true stereopsis is rare with horizontal deviation >4 PD.