On the development and comparative evaluation of an ultrasound B-mode probe calibration method
Jens Kowal1, Christoph A Amstutz, Marco Caversaccio
1M.E. Müller Research Center for Orthopaedic Surgery, Institute for Surgical Technology and Biomnechanics, University of Bern, Switzerland. jk@MEMcenter.unibe.ch
Objective:
Precise transducer calibration is an essential prerequisite for reliable surface registration based on ultrasound B-mode imaging devices. The clinical usage of a novel B-mode transducer calibration technique was evaluated and its attainable calibration precision assessed.
Materials And Methods:
The Three Wire Method and the Cambridge Calibration Method were used as reference techniques to compare the efficiency, calibration precision and spatial requirements of the different techniques. A total of 20 calibration trials were performed using each technique and were statistically evaluated for accuracy and speed.
Results:
The mean error characterizing the calibration precision of the Three Wire Method was 3.2 mm, obtained in a phantom with a volume of 14 x 10(6) mm(3) in 18.48 min. The Cambridge method resulted in a mean calibration error of 2.2 mm, but required a larger phantom with a volume of 35 x 10(6) mm3 to be used for a duration of 9.30 min. The proposed method yielded an average calibration error of 1.9 mm and was performed, on average, in 2 min using a phantom with a size smaller than 1 x 10(6) mm3.
Conclusions:
The suggested calibration method offers decreased time and space while retaining an equivalent calibration precision when compared to established reference methods.
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...