Pupil size in refractive surgery candidates.
Marcelo V Netto1, Renato Ambrósio, Steven E Wilson
1Department of Ophthalmology, University of Washington, Seattle, WA, USA.
Journal of Refractive Surgery (Thorofare, N.J. : 1995)
|August 17, 2004
Summary
Pupil size decreases with age, as measured by the Procyon pupillometer and Humphrey videokeratographer. Both devices objectively assess pupil diameter under various light conditions, with age being the primary influencing factor.
Area of Science:
- Ophthalmology
- Optometry
- Biomedical Optics
Background:
- Accurate pupil size measurement is crucial for refractive surgery planning and understanding ocular physiology.
- Variations in pupil size due to lighting conditions (scotopic, mesopic, photopic) can impact visual outcomes.
- Existing methods for pupil measurement may have limitations in accuracy or controlled conditions.
Purpose of the Study:
- To compare pupil size measurements using the Procyon pupillometer under scotopic and mesopic conditions.
- To assess pupil size measurements using the Humphrey videokeratographer under photopic conditions.
- To analyze the influence of age, gender, and refractive error on pupil size.
Main Methods:
- Pupil sizes of 96 refractive surgery candidates (192 eyes) were measured using the Procyon pupillometer PS2000 SA and Humphrey Atlas 992 corneal topographer.
- Pupil sizes were recorded under scotopic (0.04 lux), mesopic-low (0.4 lux), and mesopic-high (4 lux) illumination.
- Anisocoria and pupillary unrest were analyzed across different age groups, genders, and refractive error levels.
Main Results:
- Median pupil diameters were 6.54 mm (scotopic), 5.62 mm (mesopic-low), and 4.09 mm (mesopic-high) with the Procyon pupillometer.
- The Humphrey topographer measured a median pupil size of 3.65 mm.
- Pupil size significantly decreased with age after the fifth decade (P<.05), with no significant relationship found with gender or refractive error. Pupillary unrest was highest under mesopic-high conditions.
Conclusions:
- Both the Procyon pupillometer and Humphrey videokeratographer demonstrate an inverse correlation between pupil size and age.
- The Procyon pupillometer offers an objective and reliable method for measuring pupil size under controlled lighting.
- Age is a significant determinant of pupil size, while gender and refractive error show no significant correlation.