Marker Configuration Model-Based Roentgen Fluoroscopic Analysis.

Eric H Garling1, Bart L Kaptein, Koos Geleijns

  • 1Department of Orthopaedics, Leiden University Medical Center, P.O. Box 9600, J11-S, 2300 RC Leiden, The Netherlands. e.h.garling@lumc.nl

Journal of Biomechanics
|February 17, 2005
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

The association between tibial slope and posterior tibial tilting, not wear, in uncemented total knee arthroplasty : a secondary analysis of a five-year radiostereometry study.

The bone & joint journal·2026
Same author

Migration patterns of uncemented femoral stems in hip replacement: a systematic review and meta-analysis of clinical radiostereometric analysis cohort studies.

Acta orthopaedica·2026
Same author

The long-term revision risk of RSA-tested and non-RSA-tested TKA implants in The Netherlands: a Dutch arthroplasty register study.

Acta orthopaedica·2026
Same author

Precision evaluation of 2 CT-based radiostereometric analysis systems in a cadaver study.

Acta orthopaedica·2025
Same author

Clinical Validation and Excellent Interobserver Agreement of Volumetric Matching Micromotion Analysis (V3MA) in Total Knee Arthroplasty.

Journal of orthopaedic research : official publication of the Orthopaedic Research Society·2025
Same author

Additively manufactured shape-morphing implants for the treatment of acetabular defects.

Acta biomaterialia·2025
Same journal

Shear wave velocity of biceps femoris and medial gastrocnemius in different positions and intensities: a cross-sectional study in healthy young males.

Journal of biomechanics·2026
Same journal

Gait event detection using hybrid EMG/IMU systems: effect of SENIAM-constrained sensor placement on lower limb segments.

Journal of biomechanics·2026
Same journal

Relationship between knee adduction moment and knee contact forces during walking and running with modified foot progression angles.

Journal of biomechanics·2026
Same journal

Scaling contact force parameters across body size, limb count, and number of contact spheres.

Journal of biomechanics·2026
Same journal

The extrapolated body center of mass predicts subsequent foot placement choice during dynamic single-leg landings.

Journal of biomechanics·2026
Same journal

Lateral reactive stepping responses differ between individuals with and without transfemoral amputation.

Journal of biomechanics·2026
See all related articles

Marker Configuration Model-Based Roentgen Fluoroscopic Analysis (MCM-based RFA) accurately measures polyethylene bearing movement in knee replacements. This technique offers precise in vivo kinematic analysis using standard fluoroscopy equipment.

Area of Science:

  • Orthopedic biomechanics
  • Medical imaging analysis
  • Implant kinematics

Background:

  • Understanding polyethylene bearing movement in mobile-bearing knee replacements during dynamic activities is crucial but currently unknown.
  • Marker Configuration Model-Based Roentgen Fluoroscopic Analysis (MCM-based RFA) is a novel technique for precise implant and bone pose estimation using fluoroscopic images.

Purpose of the Study:

  • To evaluate the accuracy of MCM-based RFA in a standard fluoroscopic setup through phantom experiments.
  • To determine error propagation in MCM-based RFA using computer simulations.

Main Methods:

  • Phantom studies with a calibrated setup using tantalum markers to correct for image distortion and focus position.
  • Computer simulations to assess the influence of image distortion, model accuracy, and marker configuration on MCM-based RFA accuracy.

Related Experiment Videos

Main Results:

  • Phantom study demonstrated in-plane accuracy of 0.1 mm, out-of-plane accuracy of 0.9 mm, and rotational accuracy of 0.1 degrees.
  • Estimated worst-case in vivo accuracy: 0.14 mm (x-axis), 0.17 mm (y-axis), 1.9 mm (z-axis) translational, and 0.3 degrees rotational.

Conclusions:

  • Image distortion and model accuracy significantly impact fluoroscopic kinematic measurements.
  • MCM-based RFA shows potential as an accurate and clinically applicable tool for studying total joint replacement kinematics with standard equipment.