Background:
Our aims were to evaluate the performance of a fully automated system for measuring circulating allergen-specific IgE (sIgE) against an established in vitro assay and to assess the system's diagnostic accuracy against objective clinical criteria for identifying sensitization to specific allergens.
Methods:
Using both the IMMULITE 2000 Allergy system (IML) and an assay based on the widely used ImmunoCAP technology (CAP), we measured sIgE in serum samples from 169 persons with suspected allergies to airborne or insect venom allergens. Skin testing outcome served as the clinical comparison method.
Results:
Interassay classification agreement between the IML and CAP, relative to the usual allergen-specific IgE cutoff of 0.35 kIU/L, ranged from 76% (yellow jacket venom) to 95% (orchard grass); agreement was 88.3% for all 9 allergens combined (766 results). The 90 discordant results, when resolved by skin testing, showed better agreement with the IML (72%) than with the CAP (28%). Compared with skin testing, for each of the 9 allergens studied, the area under the ROC curve was at least as large for the IML as for the CAP, reflecting in part the more extensive working range of the IML (0.10-100 kIU/L vs 0.35-100 kIU/L for CAP).
Conclusion:
Laboratory testing for sIgE can be performed on a fully automated, random-access system with an extended working range and with diagnostic accuracy for representative allergens equivalent to or better than that of the semiautomated CAP technology.
Related Concept Videos
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or quantified.
Allergic Reactions
Cross-reactivity
Antibody Structure
Antibodies, also known as immunoglobulins (Ig), are essential players of the adaptive immune system. These antigen-binding proteins are produced by B cells and make up 20 percent of the total blood plasma by weight. In mammals, antibodies fall into five different classes, which each elicits a different biological response upon antigen binding.
The Y-Shaped Structure of Antibodies Consists of Four Polypeptide Chains
Antibodies consist of four polypeptide chains: two identical heavy...