Related Experiment Video
Updated: Aug 15, 2026

Normothermic Ex Vivo Liver Machine Perfusion in Mouse
Published on: September 25, 2023
Enhancing cell viability with pulsating flow in a hollow fiber bioartificial liver
Diem T Nguyen1, John D Brotherton, Pao C Chau
1Applied Molecular Evolutions, 3520 Dunhill St., San Diego, CA, 92121, USA.
Abstract:
A pulsating flow of medium was used to alleviate diffusion and transport limitations in a hollow fiber bioreactor containing a human hepatoblastoma cell line. The strategy is easy to implement but effective. The pulsating flow is introduced by a solenoid pinch valve at the outlet of the bioreactor and regulated by a timing circuit. In a permeability test, the system with pulsating flow had much less membrane fouling as compared to the control, a conventional hollow fiber unit. In hepatocyte culture test runs, the pulsating-flow bioreactor demonstrated the ability to maintain a higher cell viability. Histological sections indicated significantly smaller necrotic regions in the pulsating-flow bioreactor as compared to the conventional unit.