Related Experiment Video
Updated: Aug 11, 2026

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Patient dosimetry in diagnostic and interventional radiology: a practical approach using trigger levels
K Faulkner1, P Ortiz-Lopez, E Vano
1Quality Assurance Reference Centre, Wallsend, Tyne and Wear NE28 9ND, UK. keith.faulkner@nhs.net
Patient dosimetry in medical imaging faces a challenge: no single metric fully assesses both stochastic effect risks and deterministic injury likelihood. This study proposes using dose-area product (DAP) trigger levels to manage patient radiation exposure effectively.
Area of Science:
- Medical Physics
- Radiology
- Radiation Protection
Background:
- Patient dosimetry is crucial in radiology to evaluate deterministic injury risks and stochastic effect probabilities.
- A key challenge is that no single dosimetric quantity accurately assesses both risks simultaneously.
- Current practices in cardiology and interventional radiology utilize distinct dosimetry approaches.
Purpose of the Study:
- To address the limitations of current patient dosimetry methods in radiology and interventional radiology.
- To explore the utility of dose-area product (DAP) in assessing patient radiation exposure.
- To propose a framework for using DAP as a trigger for investigating potential deterministic effects.
Main Methods:
- Review of existing patient dosimetry practices in radiology and interventional radiology.
- Analysis of the correlation between effective dose and dose-area product (DAP).
- Evaluation of DAP's correlation with maximum skin dose for predicting deterministic injuries.
Main Results:
- Effective dose correlates well with stochastic effect risk but not with maximum skin dose.
- Dose-area product (DAP) does not correlate with maximum skin dose, a predictor of deterministic injuries.
- DAP readings can serve as effective trigger levels for investigating maximum skin entrance dose.
Conclusions:
- A single quantity cannot adequately assess both stochastic and deterministic risks in patient dosimetry.
- Dose-area product (DAP) is valuable for monitoring and managing patient radiation exposure in interventional procedures.
- Proposed trigger levels for DAP can enhance patient safety by prompting investigations into potential skin injuries.
More Related Videos
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Radiological Investigation I: X-ray and CT
Biological Effects of Radiation
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...